Characterization of a HPGe detector response for activation cross section measurements: regression method versus Monte Carlo method
Description
The full energy peak efficiencies and covariances of a high purity germanium (HPGe) detector
studied by the regression method and Monte Carlo method were compared. The impact of the
obtained results on the neutron activation cross-sections measured relative to monitor crosssections
was studied. In the regression method, the efficiencies measured for a set of calibration
point sources were analysed by the least-squares analysis. In the Monte Carlo method, the
efficiencies for the calibration point sources were calculated by MCNP. The covariances of the
efficiencies determined by the regression method were calculated analytically. Perturbation
analysis was performed to estimate the covariances of the efficiencies calculated by the Monte
Carlo method. Positive correlations higher than 0.8 were found in the uncertainties of the MCNP
data for point-like sources. In the case of the regression method, the correlation coefficient
contains both positive and negative terms. The efficiencies and their covariances for finite sample
geometry were also estimated by both methods to take into account sample effects such as
geometrical effect, and gamma-ray self-absorption, and found considerable differences in the
cross-sections and their uncertainties for reaction products quantified with low energy gammas.
The efficiencies and covariances were clearly affected by the properties of the sample.